• image
  • image
  • image
  • image
  • image

Rotofix 46

ROTOFIX 46 :
The ROTOFIX 46 is the ideal centrifuge for research and industrial laboratories. It has a wide range of accessories for routine applications and also enables special tasks to be tackled.

This centrifuge can accommodate a number of special tubes, including ASTM (American Society for Testing Materials) tubes, Schlenk tubes and chrome bath tubes.

Rotor 5616 can hold Schlenk tubes, which are widely used in organometallic and co-ordination chemistry. Substances that have been prepared in Schlenk tubes can be separated through centrifugation.

The sulphuric acid content of chrome baths can be determined using chrome bath tubes. The QA procedure is of interest not only to public and private test facilities, but also to operators of chrome baths who need to establish sulphate levels at regular intervals.

The ROTOFIX 46 is not a medicinal product within the meaning of the IVD Directive and therefore must not be used for medical analysis / diagnosis. The IVD Directive is valid throughout Europe and in other countries that have adopted it or use it as a basis for their legislation.

PRODUCT DETAILS

Max. capacity in a swing-out rotor:4 x 290 ml
Max. speed (RPM):4,000
Max. RCF:3,095
Dimensions (H x W x D):345 x 538 x 647 mm
Weight:approx. 60 kg
Refrigeration:Air cooling
Frequently used rotors:Angle rotor for 6 x 25 or 50 ml Schlenk tubes (45°), max. RCF 805 – Cat. No. 5616
Swing-out rotor for 6 x 30 ml chrome bath tubes (90°), max. RCF 917 – Cat. No. 4619
Swing-out rotor for max. 4 x 290 ml (90°), max. RCF 3,095 – Cat. No. 5694
As their designated purpose is for general laboratory use only, the ROTOFIX 46 and 46 H are not medical products according to the IvD Directive. These centrifuges must, therefore, not be used for medical analyses or diagnostic purposes (clinical chemistry) within the scope of the IvD Directive.

OPERATOR COMFORT
The ROTOFIX 46 offers various features to make operation easier, such as  
Rotary opening mechanism so that the centrifuge can be easily opened and closed
Ergonomic operating and information panel
10 programmable memories
The parameter combination last selected is retained after the centrifuge is switched off
After entry of the rotor radius r in mm it is possible to switch between display of rpm and the relative centrifugal force (RCF) by pressing the RCF button
The parameters can be rapidly entered and changed using the rotary button

SAFETY
The ROTOFIX 46 contributes to safety in the workplace. This centrifuge offers the following safety features:
Lid locking and latching mechanism
Automatic rotor recognition
Imbalance switch-off
Emergency lid release
Lid dropping prevention mechanism
Motor overheating protection
Chamber overheating protection



DESIGN
This compact centrifuge has attractive design features such as:
Metal casing
Metal lid
Centrifugation chamber made of rust-free stainless steel
View port in the lid

  • PDF Files Download
  • Rotofix_46_02130_2202.pdf 146KB